10-20 ADS FlowShark Manual

Problem

No DC power at EMU.

Possible

Cause

Breaker has tripped.

Station power is off.

Power supply has failed.

Power supply wiring has become disconnected.

Possible

Solutions

Restore main power source to station (have owner
perform this).

Check for wiring faults and then reset breaker.

Look for “DC ON” LED on front of ADS Power Supply
module. If off, but AC power is live, replace power
supply.

Contact your regional ADS representative.

Ultrasonic Depth Subsystem

The following tables contain general troubleshooting techniques for
problems with the ultrasonic depth subsystem.

Note: Contact your regional ADS representative for
further diagnosis prior to replacing a sensor.

Problem

Range from ultrasonic depth subsystem is slightly
inconsistent with manually measured range.

Possible

Causes

Temperature or physical offset(s) may be incorrect.

Pipe height may be incorrect.

Possible

Solutions

Verify the temperature and physical offsets, and adjust
offsets if necessary.

Verify the pipe height.