Pressure depth subsystem – ADS Environmental Services FlowShark QR 775002 A4 User Manual

Page 237

Advertising
background image

Maintenance and Troubleshooting 10-25

Problem

Velocity data seems erratic.

Possible

Causes

Velocity sensor may be dirty.

Velocity sensor may be faulty.

Analog board may be faulty.

Possible

Solutions

Clean the velocity sensor.

Check the sensor connections.

Check the velocity parameters.

Contact your regional ADS representative.

Replace the velocity sensor.

Replace the monitor if defective.

Problem

Velocity readings are abnormally high.

Possible

Causes

Doppler velocity sensor may be covered with silt.

Velocity sensor may be out of the flow.

Flow may be reversed or slower than 0.5 ft/s (0.15 m/s).

Possible

Solutions

Check the site conditions, and relocate the velocity
sensor if necessary.

Pressure Depth Subsystem

The following tables contain general troubleshooting techniques for
the pressure depth subsystem.

Warning: Contact your regional ADS representative for
further diagnosis before replacing a sensor.