10-24 ADS FlowShark Manual

Problem

Monitor often provides a velocity reading of 0.

Possible

Causes

Doppler velocity sensor may be dirty or broken.

Cabling between the analog board and the velocity
sensor may be bad.

Velocity parameters may require adjustment.

Possible

Solutions

Clean the velocity sensor.

Check the velocity parameters.

Inspect the velocity sensor cables for tightness.

Check for moisture in the connector.

Contact your regional ADS representative.

Replace the velocity sensor.

Replace the monitor if defective.

Problem

Velocity data does not fluctuate much (but is not 0).

Possible

Cause

Doppler velocity sensor may be dirty

Velocity sensor may be broken.

Possible

Solutions

Clean the velocity sensor.

Check the velocity sensor connections.

Check the velocity parameters.

Contact your regional ADS representative.

Replace the velocity sensor.