Analog Input/Output Installation and Configuration 8-3

Analog Inputs

FlowShark models 5000-CG, 5000-CG-1, 5000-CL, and 5000-CS
support the monitoring of two analog input 4-20mA signals from
other instruments. These signals typically represent measurements
made by other devices, such as flow meters, pressure sensors, or
level transducers. To properly configure the monitor using Profile

®

,

you must know the signal scale and engineering units (e.g., 4 to
20mA = 0 to 20MGD).

Perform the following steps to connect analog inputs (maximum of
two) to the ADS FlowShark monitor.

Gathering Tools and Equipment

Gather the following tools and materials in preparation for analog
input connection:

Laptop with Profile 3.4 or greater

LIF configured with analog parameters, sample rate, etc.

PC comm cable (ADS p/n 5000-0601)

Analog I/O cable (ADS p/n 5000-0698)

Electrical tape

Cable ties

Cable grip fitting with gland for .4” cable

Hole saw or punch to fit cable gland

Pilot drill if punch used

Drill

Vacuum (to remove cuttings from box if hole saw used)

Hand tools (strippers, screwdrivers, wrenches, etc.)

Digital multimeter