102 | S3 and S3e Cell Sorters

Maintenance

Fig. 122. Adjust Stream Position window.

29. Using the Pitch and Roll arrows, adjust the stream’s position until the stream is hitting the

waste hole.

30. Click the blue forward arrow.
31. Close the sort chamber door.
32. Using the Pitch and Roll arrows, move the stream until it is visible as a vertical white line in

the camera image on the right of the window (Figure 12).

33. Adjust the stream so that the red dotted lines become green.
34. Click the blue forward arrow.
Note: When the stream is in line with the stream laser, the illuminated stream will become
thicker and more visible.

35. Using the Vertical, Horizontal, and Focus arrows, adjust the nozzle tip and stream until it

matches the last aligned image (Figure 123).

36. If there appears to be a clog or if the stream appears to be uneven, repeat cleaning the tip

and reseating the nozzle as described starting on step 1.

37. If the nozzle tip and stream are aligned and there are no clogs, click the checkmark to

proceed (Figure 123).