Bio-Rad S3™ Cell Sorter User Manual
Page 113
Advertising
Instruction
Manual | 103
10.4 Swap Nozzle Tip Wizard
38. The AutoGimbal system will commence automatic alignment of the nozzle tip and stream
to the optics.
During the automatic alignment process, a series of windows will appear (Figure 124). This
process can take 5–6 min to complete.
Fig. 123. Adjust Stream Position window.
Fig. 124. Automated alignment process windows.
Advertising
This manual is related to the following products: