Troubleshooting, Troubleshooting 111 – Bio-Rad S3™ Cell Sorter User Manual

Page 121

Advertising
background image

Instruction

Manual | 111

Troubleshooting

11

Table 30 describes possible errors and causes with troubleshooting steps.

Table 30. Troubleshooting table.
Error

Possible Causes

Troubleshooting Steps

Bubble detected

Bubble in sample line

Place the loading dock in the wash position
to backflush the sample line and remove
the bubble

Nozzle door interlock

Nozzle door is open

Close nozzle door

Sort door interlock

Sort door is open

Close sort door

Waiting for QC

QC has not been run since the

Run QC to proceed with sorting

system has been started up

Not maintaining droplets

Large shift in drops

Perform Drop Delay to put the system back
into maintaining droplets

Cannot unlock loading stage

System is still washing

Wait for wash to complete

Sample is currently running

Stop sort or acquisition

QC will not pass

Calibration beads diluted

Use the undiluted ProLine

beads directly

from bottle

Using wrong calibration beads

Use only ProLine beads

QC criteria changed

Refer to the correct bead lot file

Fluidic error message

Quick disconnects are not

Recheck connections. An audible click must

connected properly

be heard for proper connection

Stream coming out at an angle

Clog in nozzle tip

Use Swap Tip Wizard and clean nozzle tip

Reseat nozzle tip on to the nozzle

Manual compensation

Compensation checkbox

Check and uncheck compensation

not working

not checked

box on the axis

Cannot see events labeled

Filter configuration may

Check filter configuration against

with fluorophores

be incorrect

fluorophore combination

Laser turned off

Check that both lasers are powered

Cannot see rare events

Total events displayed is

Increase cycle mode value to display

too low

more events

continues