7.5 Instrument Status Box

The cleaning protocol will be completed when the window disappears. If a high pressure
cleaning has been chosen, run a QC using ProLine

calibration beads before sorting.

7.5 Instrument Status Box

The instrument status box is located at the bottom of the instrument control panel. The
notification and the color of the box will change depending on the state of the instrument. Table
26 describes the notifications and colors.

Table 26. Instrument Status Box notifications and colors.
Color

Notification

Action

Example

Gray

Off

Nothing

Off

Green

Ready

Can begin acquisition or sort

Acquiring sorting

Yellow

Information

Perform requested action

Require QC

before proceeding

Fluids Low/High

Orange

Warning

Nothing, system is running

Calibrating

Red

Error

Perform required maintenance

Not maintaining droplets

7.6 Status Bar

The status bar is located at the bottom of the ProSort Software window with nine status
indicators (Figure 82). Table 27 describes the status indicators.

Table 27. Description of each status indicator.
Number

Description

1

Software connectivity to instrument

2

Current date

3

Temperature control status

4

Event rate (events per second)

5

Total event number

6

Acquisition time elapsed

7

Sort time elapsed

8

Current user

Fig. 82. Status bar with the status indicators numbered.

Fig. 81. Cleaning protocol final stage.