3.2 Control Panel

Table 20. Instrument controls and their functions.
Button

Name

Function

Start Sample

Starts sample acquisition. If a data file is loaded and it has been saved,

Acquisition

the data will clear and acquisition will start

Stop Sample

Stops acquisition and sample flow. Regions, gates, and annotations may

Acquisition

be modified before saving the FCS file

Start Sorting

Starts sorting. If a data file is loaded and it has been saved, the data will
clear and sorting will start

Stop Sorting

Stops sorting and sample flow. Regions, gates, and annotations may be
modified before saving the FCS file

Clear Acquired Data Clears acquired data. All plots and statistics will be refreshed.

Pause Sample

Pauses sample acquisition and sample flow. The data at this point may be
appended (by the Resume function) or saved as is. Saving after pausing will
disable the Resume function. To remove sample tube, unlock the touchscreen
locking system located above the loading stage after pausing acquisition

Resume Sample

Resumes acquisition when a sample has been paused. Resuming will
append the paused FCS file and the time paused can be seen on a time
histogram. If you have removed the sample tube it must be returned and
relocked before acquisition can be resumed

Eject Sample Tube

Depressurizes and unlocks the sample. The loading stage can be moved
into wash position to access sample. The touchscreen locking system can
also be used to unlock the sample

Adjust Sample

Starts or stops agitation of the sample. There are three levels of agitation:

Agitation

low, high, and off. Click this button once to set to low, twice to set to high,
and three times to turn off agitation

Auto Save

When checked, automatically prompted to save a data file after
acquisition stops.

Target Event Rate

Allows selection of target event rate for sample acquisition or use of preset

sample rates low/med/high based on sample concentration

Limit

When checked, stops data acquisition after the number of events entered
has been reached

Gate Limit

When checked, data acquisition stops after the number of events entered
has been reached within the selected gate

Cycle Mode

When checked, automatically refreshes the data displayed according to
the entered time. The acquired data set continues to accumulate in the file

Sample Chamber

Turns the sample chamber light on and off

Light

Enable/Disable Sort Illuminates entire sort collection chamber for easy viewing of deflection

Chamber Light

plates, sort stream, and collection vessels

Enable/Disable

Enables or disables temperature control of the sample station and sort

Temperature Control collection vessel

Biosafety

Indicates the status of the biosafety system and gives information on filter life.

System Status

If the biosafety system loses connection, cannot connect, or if the filter life is
at 0% an error will occur and not allow the user to sort or acquire.

continues