Bio-Rad S3™ Cell Sorter User Manual

Page 122

Advertising
background image

112 | S3 and S3e Cell Sorters

Troubleshooting

Table 30. Troubleshooting table (continued).
Error

Possible Causes

Troubleshooting Steps

Drastic light change

Background movement in

Redo QC and avoid changing light

front of green sliding door

conditions in front of green sliding door

during QC process

during the procedure

Sheath error

The internal sheath

The connections to the DI water and 8x

reservoir cannot maintain

sheath fluid containers should be checked to

proper levels

make sure they are connected, that they
are not empty, and that fluid can flow. It is
possible that their filters are clogged and
need to be replaced

Waste error

Waste quick disconnect

The connections to the waste container

not connected properly to

should be checked to make sure they are

waste container

connected and that waste can flow. Once

corrected, it will take time for the system to

Internal waste reservoir filled

clear the internal reservoir. You can observe
that pumping is happening by watching for
splashing in the waste container

Nozzle stage cannot return to

Laser intersection cover has

Adjust laser intersection cover so it is seated

proper position after nozzle

been moved during tip swap

correctly. Once in the proper place the

tip swap

nozzle stage can be lowered into position

Cannot acquire sample.

Room too cold or faulty

Warm room to >63°F. If this does not help,

Hovering on acquisition button

bubble detector

contact service support

displays “bubble detected”

Arcing

The high voltage deflection

Stream spray or droplets are on the

plate monitor has measured

deflection plates. Carefully clean and dry

the plate voltage deviating

the plates

from normal

No data appear on the work-

Nozzle interlock has been

Open nozzle door all the way then close to

space after changing nozzle tip

activated

reset the laser interlock

No events during acquisition

Nozzle door open

Close the nozzle door

Laser shutter closed

Cycle the nozzle door open then closed,
especially if cover was removed

Laser not on

Turn on the laser in the user interface
Restart instrument and computer to correct
bad communication

Threshold too high

Lower threshold

Triggering on noise

Change the trigger parameter to FSC or SSC

Stream not centered in

Use the electronic micromotors (Agillis)

laser beam

positioner while acquiring to maximize event
signal — access within the system log file

No beads in sample

Check sample tube, vortex

continues