6.2 Sort Modes

6.2 Sort Modes

Table 24 describes the three sort modes available when setting up a sort. Sort modes can be
selected for each sort direction.

Table 24. Sort mode definitions.
Sort Mode

Definition

When to Use

Single Cell

This mode requires that one and only one positive event be in the

Single cell deposition

center of the droplet without any negative events nearby. Recovery
will be reduced due to the sort restrictions

Purity

This mode will sort as long as no negative events are present in the

General sorts where both purity

droplet. If multiple positive events are present, sorting will still occur

and recovery are important

Enrich

This mode will sort all positive events regardless if negative events

Sorts when recovery is critical

are nearby. Purity will be reduced to maintain high recovery

and purity is not as important

6.3 Sort Statistics

The sort statistics window can be viewed during
a sort and added to a print page for reporting
(Figure 75). This window will automatically open
when a sort is started.

If a saved partial FCS file is opened, aggregate
statistics can be shown:
1. File > Open FCS File with Protocol.
2. Click on the Home tab and click Sort

Statistics.

3. Check the Show Aggregate Statistics box

to view statistics from all sorted events.

Fig. 74. Sort Logic window with sort priority and sort modes highlighted.

Fig. 75. Sort Statistics window.