114 | S3 and S3e Cell Sorters

Troubleshooting

Table 30. Troubleshooting table (continued).
Error

Possible Causes

Troubleshooting Steps

Decontamination procedure

Software error

Contact technical support for help on the phone.

freezes during priming

Stream is spraying everywhere No tip in place

Run nozzle tip swap wizard and install nozzle.

No O-ring in place

Run nozzle tip swap wizard to reseat nozzle
with O-ring.

Tip is still clogged

Run nozzle tip swap wizard to clean nozzle using
sonication and syringe wash. Reseat nozzle.

Stream is spraying at an

Tip was not seated straight

Run nozzle tip swap wizard to clean nozzle and

angle with spray.

then reseat nozzle correctly.

No O-ring in place

Run nozzle tip swap wizard to reseat nozzle
with O-ring.

Wagon wheel has fallen out,

Run nozzle tip swap wizard to check if wagon

extra wagon wheel installed,

wheel is installed properly. Reseat nozzle.

or wagon wheel installed

incorrectly.

Tip is still clogged

Run nozzle tip swap wizard to clean nozzle using
sonication and syringe wash. Reseat nozzle.

Stream is moving around,

No O-ring in place

Run nozzle tip swap wizard to reseat nozzle

droplet camera shows

with O-ring.

droplets are moving

Wagon wheel has fallen out,

Run nozzle tip swap wizard to check if wagon

inconsistently

extra wagon wheel installed,

wheel is installed properly. Reseat nozzle.

or wagon wheel installed

incorrectly.

Tip is still clogged

Run nozzle tip swap wizard to clean nozzle using
sonication and syringe wash. Reseat nozzle.

Stream is hitting the back

Tip was not seated straight

Run nozzle tip swap wizard to clean nozzle and then

wall, deflection plates or

reseat nozzle correctly.

sort tray

No O-ring in place

Run nozzle tip swap wizard to reseat nozzle
with O-ring.

Wagon wheel has fallen out,

Run nozzle tip swap wizard to check check if wagon

extra wagon wheel, or wagon

wheel is installed properly. Reseat nozzle.

wheel installed incorrectly.

Tip is still clogged

Run nozzle tip swap wizard to clean nozzle using
sonication and syringe wash. Reseat nozzle.