Bio-Rad S3™ Cell Sorter User Manual

Page 123

Advertising
background image

Instruction

Manual | 113

Troubleshooting

Table 30. Troubleshooting table (continued).
Error

Possible Causes

Troubleshooting Steps

High event rate (while trying

Drop drive noise causing

Toggle between the FSC/SSC trigger (in the

to run QC, or running a

excessive event rate

service panel)

sample at 500 eps)

(>40,000 eps)

Perform a nozzle tip swap

Calibrate droplets (located in the Edit Droplets

window when logged in as administrator)

Too much light. Event rate

Light leak in system, ensure all covers are closed

(~8,000 eps)

Check that the filter blocks or filters have not been
removed or that they are seated in proper position

If triggering on FSC, may be misaligned or ND

filter removed

Obscuration bar may need adjustment. Contact
Bio-Rad Technical Support

Inappropriate sample pressure. Perform Calibrate Sample Pressure Offset
Event rate (~2,000 eps)

Running samples at above

Too much light

Obscuration bar may be flat and may need

2,000 eps and getting

adjustment. Contact Bio-Rad Technical Support

40,000 eps

Threshold may be set too low, raise threshold value

Signal intensity is too bright. Add ND filter

Poor drop delay (uneven V,

Bad side streams or center

Set defanning manually until the center stream is

flat on one side)

stream

solid and tight

Adjust phase to ensure the side streams are tight

Recalibrate droplets located in Droplet Control

window on the admin tab or at the end of a nozzle
tip swap procedure

Side stream deflection not set properly. Adjust side
stream deflection on the admin tab

Very little volume is used

Pinch in the sample line

If sample line is partially clogged, run clean cycle

while performing cleaning

multiple times with backflushing to remove clog

user logout or prior to
(after system shutdown)
Uses 2 ml of 10% bleach/
water takes approx. 1 min
Usually 2 ml of 70% ethanol
takes approx. 2 min

Waste Trough Filling, System Waste trough port clogged

Clear debris from waste trough port and rinse with

goes into Safe Mode

and overfilling

bleach. Restart instrument and computer.

Trough is empty, droplet

Check to see if trough has liquid or if waste trough

trapped on sensor that causes

port is clogged. If not use canned air to clear

sensor to trip indiscriminately

droplet form sensor as a temporary fix. Restart
instrument and computer. Contact service support
for repair or replacement.

Variable event rate during

Agitator seal may be leaky

Turn off the agitator to see if it resolves the varying

acquisition (changes at

event rate. If it does, contact service support for

regular intervals)

replacement.

Possible nozzle clog

Clean nozzle tip

Hardware issue

Contact Bio-Rad Technical Support

Waste container is

Hydrophobic filter is wet

Contact service support for replacement.

abnormally expanded

continues

If only a very small amount of liquid (200–300 µl)
is taken from the tube containing the cleaning
solution during a cleaning protocol, the sample
line may be pinched, and may need replacement.
Sometimes massaging the pinched area in the line
(if visible) may resolve the issue. Contact Bio-Rad
Technical Support