2 | S3 and S3e Cell Sorters

Introduction

The S3 or S3e

Cell Sorter complete system includes the following components:

n

S3 or S3e Cell Sorter Instrument, 1 each

n

S3 or S3e Installation Kit, 1 each

S3 cap assembly, white, sheath fluid, 1 each
S3 cap assembly, blue, DI water, 1 each
S3 cap assembly, red, waste, 1 each
S3 fluidics container 4 L size, empty, 2 each
S3 individual filter holder, black, empty, 5 each
S3 forward scatter (FSC) filter holder
with neutral density filter, 2.0, 1 each
S3 filter block A with preset filter
configuration, 1 each
S3 filter block B with preset filter
configuration, 1 each

5 ml tubes, 12 x 75 mm, 1 pack
S3 collection adaptor set, 1 each
Bulk fluidics tray, 3 each
USB cord, 1 each
Ethernet cord, 1 each
Power cord, 1 each
USB drive with ProSort

software

Instrument and software manual
Instrument quick guide

n

S3 Accessory Kit (optional), 1 each (catalog #145-1065)
S3 nozzle tip, 100 µm, 1 each
S3 nozzle O-rings, 2 each
S3 nozzle alignment disk, 2 each
Neutral density filter, 1.0, 2 each

2 mm hex driver, 1 each
S3 spanner wrench, 1 each
1 ml syringe, 1 each
Plastic box, 1 each

n

Computer CPU and 24 inch monitor, 1 each (catalog #145-1066)

n

ProFlow

8x Sheath Fluid, preservative free, 5 x 4 L, 1 case (catalog #145-1082)

n

ProLine

Calibration Beads, 3 x 5 ml, 1 pack (catalog #145-1081) or

ProLine Universal Calibration Beads, 3 x 5 ml, 1 pack (catalog #145-1086)

Table 5. Dimensions for instrument alone or with computer and monitor.*
Instrument only (W x D x H)

70 x 65 x 65 cm

27.5 x 25.5 x 25.5 in

Instrument wiith computer and monitor,

116 x 65 x 65 cm

(W x D x H)

46 x 25.5 x 25.5 in

* An additional 61 cm (24 in) of height clearance is needed for service.

1.1 System Components

1.2 Installation Requirements

The S3 and S3e

Cell Sorters should be installed by a trained service engineer to ensure proper

operation and calibration of the instrument. If any items are missing or damaged, contact your
local Bio-Rad office for assistance.

Before the S3 and S3e

Cell Sorters can be installed by a service engineer, a site must be

chosen. The instrument should be located on a sturdy bench or table top, away from any other
instruments that may interfere electrically or mechanically by causing vibration. The bench or
table top must be able to accommodate 198 lb (90 kg), the weight of the instrument. The area
should be free of excessive dust or moisture.