Bio-Rad S3™ Cell Sorter User Manual

Page 91

Advertising
background image

Instruction

Manual | 81

8.1 Daily Shutdown

A window will open requesting a tube of cleaner to be used to clean the system (Figure 91).

6. Load a tube of cleaning solution.
7. Move the loading stage into the run position.

8. The system will automatically run through the cleaning procedure, which takes

approximately 2 min, while displaying status windows (Figure 92).

9. Put the Loading Stage into the wash position.

Fig. 91. Start the clean system option by following the prompt to load a tube of cleaner.

Fig. 92. Cleaning status windows.

Advertising
This manual is related to the following products: