Report from completed calibration processes – KERN & SOHN AET 200-4NM User Manual

Page 101

Advertising
background image

AET_/PET_/ILT_NM-BA-e-1410

101

Follow below pipettes calibration procedure:

1. Press <

Select pipette> key;

2. Select a pipette for testing from displayed list;

3. On selecting, the software returns to the main window, and name of

selected pipette is visible in the grey coloured workspace;

4. Press <

Start> key in the bottom bar of the display;

5. If option <Request batch number> is enabled (set to <YES>) a window is

opened for inserting the batch number, and accepting the value;

6. A window is opened for inserting values of ambient conditions parameters

read from the external sensors (temperature, humidity, pressure and

water temperature), on inserting press <

Accept> key;

If option <Acquire ambient conditions from THB module> is enabled (set

to <YES>, the software automatically acquires data from the THB module

on temperature, humidity and atmospheric pressure. The value of water

temperature has to be inserted manually, and confirmed by pressing <

Accept> key;

7. The software returns to the main window, and the grey coloured

workspace displays the following data: tested volume and process status

<In progress>. The task bar displays commands and hits for a user on the

following step - <Weighing sample C1/V1/N1>. Description of texts: C1 –

channel number; V1 – number of volume for a channel; N1 – number of

measurement for tested volume;

8. Follow displayed commands until completing the calibration process. The

grey coloured workspace displays other data on calibration process, i.e.

average volume, values of errors, specific for carried out process;

9. On accepting the last measurement, a window is displayed with ambient

conditions parameters. Proceed as on process start and insert all ambient

conditions parameters, and press <

Accept> key;

10. The software automatically generates a report, that is printed and stored

in the database of reports from pipettes calibration (ambient conditions

and water temperature values given in a report are averages from the

initial and final values of a process);

11. Complete the process by pressing <

> key visible in the bottom bar.

In case of multichannel pipettes, the balance displays a message on

continuing the process for the following channels. On confirming the

message, the balance carries out calibration / testing process for the

following channel. The settings remain unchanged;

12. The software enables starting the following procedure for the same

pipette or selecting another pipette from the database.

23.6. Report from completed calibration processes

On completing each calibration process, a report is generated. The report is

saved in a database <

Reports from pipettes calibration

processes>. The files are named by their date and time (hour) of

calibration process.