AET_/PET_/ILT_NM-BA-e-1410

AUTOTEST FILTER

It is process of 10 times loading and unloading of the internal weight for

all possible settings of balance’s filter and value release values. The test

controls 2 parameters:

Repeatability;

Stabilization time of measurement result.

The whole testing process takes approximately 1 hour. On completing the

test for all possible settings, balance displays a message box with

obtained results.

The user is given information on optimum balance settings in present

ambient conditions at a workstation. The function provides a user with

data on the best possible balance settings for obtaining the highest

measurement repeatability and the shortest measurement time with

acceptable value of repeatability.

Test results are not stored in balance’s memory.

Exiting function causes erasing the test results.

Additionally the function enables:

Printing test results on printers available in the system and selecting

the most optimum parameters from the balance’s option level.

Measurement results:

*Filter type

*Setting of “Value release” parameter

*Repeatability of indications expressed as standard deviation

*Average time of measurement stabilization