Settings:

VALUE – selecting this option opens a window for inserting

determined value of air density (e.g.: determined using other

methods). The inserted value is used to calculating buoyancy

compensation factor. The value is assigned automatically on

carrying out the air density determination (on its completion by
pressing <

> key).

ONLINE – selecting this option downloads current air density value

from a connected THB sensor, if such is connected to the balance,

or from internal sensor installed in the balance.

If a balance features both types of sensors (THB module and

internal sensor), then data from the THB module supersedes the

other, and parameters Setup/Ambient Conditions/Ambient

Conditions Module require setting to SAVE or SAVE & ALERTS

mode.

If a balance features only one type of the sensors for temperature,

humidity and pressure recording, then correct ONLINE operation

requires setting of balance parameters:

Internal sensors only - Setup/Ambient Conditions/Ambient

Conditions Module set to NONE,

External THB module only - Setup/Ambient

Conditions/Ambient Conditions Module set to SAVE or

SAVE & ALERTS mode.

Additionally, for correct balance cooperation with the THB module

set port’s transmission parameters to compatible with those of the

THB module. the THB module transmission parameters are detailed

on its data plate.

On returning to the weighing mode, the display indicates an additional symbol
<

!

>, as indicated on below drawing.

From now on the value of indicated mass

is corrected including the air buoyancy compensation factor and sample

density.

The measurement result can only be compensated by a correct value if an

operator inserts to balance’s memory the actual value on the air density and

correct density of weighed sample.