KERN & SOHN AET 200-4NM User Manual

Page 33

Advertising
background image

AET_/PET_/ILT_NM-BA-e-1410

33

7. Which opens a window containing a list of minimum sample methods

stored in balance database.

8. Select one of available entries.

9. The software returns to the previous menu window

10. Press <Mode> field

11. Which opens a window with parameter settings. Select one of below

options:

Lock – on selecting this option and during the weighing process the

display indicates corresponding pictograms that inform a user whether

the weighed mass is below or above the minimum sample limit.

Additionally the software disables accepting the measurement result if it

is below the minimum sample limit

Warn – on selecting this option and during the weighing process the

display indicates corresponding pictograms that inform a user whether

the weighed mass is below or above the minimum sample limit. The

user can accept a measurement that is below the minimum sample

limit, but on a printout such measurement is preceded by an asterisk

(*).

12. On setting the parameters exit the menu.

13. The mass field on the display is supplemented by an additional

informative pictogram. The pictogram changes during weighing process

and indicates current status of mass in relation to the pre-determined

value of minimum sample.

- Mass below the set value of minimum

sample

- Mass above the set value of minimum

sample for a given tare range