background image

AET_/PET_/ILT_NM-BA-e-1410

165

Frame format:

1 2-3

4

5

6

7-15

16 17 18 19 20 21

S space stability

marker space character mass space

unit

CR LF

Example:
S CR LF
– command sent from a computer

S _ A CR LF - command understood and in progress

S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF - command carried out, response

is mass value in basic measuring unit.

Immediately send measurement result in basic measuring unit

Format: SI CR LF

Accessible responses:

SI_I CR LF

- command understood but not accessible at this

moment

MASS FRAME

- response is immediate with mass value in basic

weighing unit

Frame format:

1 2

3

4

5

6

7-15

16 17 18 19 20 21

S I space stability

marker space character mass space

unit

CR LF

Example:
S I CR LF
– command sent from a computer

S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command carried out,

immediate response of mass value in basic weighing unit

Immediately send measurement results from all weighing platforms

basic measuring units

Format: SIA CR LF

Accessible responses:

SIA_I CR LF

- command understood but not accessible at

this moment

MASS FRAME „P1” CR LF

MASS FRAME „P2” CR LF

- immediate response of mass value in basic

measuring unit from all weighing platforms

Frame format with mass from each weighing platform:

1 2

3

4

5

6

7-15

16 17 18 19 20 21

P n space stability

marker space character mass space

unit

CR LF

Advertising