User adjustment, Adjustment test, User adjustment 10.4. adjustment test – KERN & SOHN AET 200-4NM User Manual

Page 36

Advertising
background image

36

AET_/PET_/ILT_NM-BA-e-1410

Acting according to the command, place an ordered weight/standard on

balance’s weighing pan and press

key,

On completing the adjustment procedure the balance indicates a message

box:

On confirming the message box with

key, the balance returns to

weighing mode.

10.3. User adjustment

User adjustment is carried out with an optional standard, which mass ranges

between 0,15 Max and Max. Adjustment procedure is compatible with the

external adjustment process, but before its start the software opens a

message box for entering mass of a standard used for user adjustment.

CAUTION:

User adjustment is available only in balances which are not subject to conformity

assessment (verification).

In order to start user adjustment, enter submenu < Adjustment >, and select
option: “ User adjustment”. Then follow the commands indicated on

balance’s display.

10.4. Adjustment test

< Adjustment test > function enables comparing the result of internal automatic

adjustment with the value of internal weight saved in balance’s factory

parameters. The comparison is used for determining balance’s sensitivity drift

over time.