KERN & SOHN AET 200-4NM User Manual

Page 163

Advertising
background image

AET_/PET_/ILT_NM-BA-e-1410

163

33.2. Responses format for commands sent from computer level

On receipt of a command, the terminal sends a response in one of the

following formats:

XX_A CR LF

command understood and in progress

XX_D CR LF

command carried out (appears only after the

command XX_A)

XX_I CR LF

command understood but not accessible at this

moment

XX _ ^ CR LF

command understood but max range is exceeded

XX _ v CR LF

command understood but min range is exceeded

XX _ OK CR LF

Command carried out (completed)

ES_CR LF

Command not recognized

XX _ E CR LF

an error occurred on command carrying out (time

limit exceeded while waiting for stable measurement

result (time limit is balance’s characteristic

parameter)

XX - stands for a name of sent command

_ - substitutes spaces

COMMANDS DESCRIPTION

Zero the balance

Format: Z CR LF

Accessible responses:

Z_A CR LF

Z_D CR LF

- command understood and in progress

- command carried out

Z_A CR LF

Z_^ CR LF

- command understood and in progress

- command understood but zeroing range exceeded

Z_A CR LF

Z_E CR LF

- command understood and in progress

- time limit exceeded while waiting for stable measurement

result

Z_I CR LF

- command understood but not accessible at this moment