Report from product control – KERN & SOHN AET 200-4NM User Manual

Page 122

Advertising
background image

122

AET_/PET_/ILT_NM-BA-e-1410

25.4. Report from product control

An example of a report

:

----------------- SQC ------------------

User

Product 000

Start date 2012.10.19 12:27:09

End date 2012.10.19 12:37:20

Batch number ppp

Batch quantity 10

Nominal mass 50 g

Limit T2- 5 g 10 %

Limit T1- 2.5 g 5 %

Limit T1+ 1 g 2 %

Limit T2+ 2 g 4 %

--------- Measurement 1 --------------

Net 49.998 g

--------- Measurement 2 --------------

Net 49.998 g

--------- Measurement 3 --------------

Net 50.021 g

---------Measurement 4 ---------------

Net 50.038 g

---------Measurement 5 ---------------

Net 50.053 g

---------Measurement 6 ---------------

Net 50.030 g

---------Measurement 7 ---------------

Net 50.012 g

---------Measurement 8 ---------------

Net 50.029 g

---------Measurement 9 ---------------

Net 50.007 g

---------Measurement 10 -------------

Net 50.008 g

Number of T2- errors 0 0 %

Number of T1- errors 0 0 %

Number of T1+ errors 0 0 %

Number of T2+ errors 0 0 %

Average 50.0194 g

Standard deviation 0.01714 g

----------------------------------------

Signature

........................................

Report template:
The balance user in the submenu <

Printouts /

Printout template of a

SQC report > can edit a template of a report from product control. The

variables attributed to <YES> are printed in a report.