PROCEDURE:

On function activation, the software initiates the process automatically. The

display indicates progress of the process course. On completing the process, the

software displays a summary and marked values of currently set filter values.

The report can be printed.

Caution:

Exiting the reporting window causes erasing the test results.

User can abort the process at optional moment by pressing <X> key in the

testing window.

AUTOTEST GLP

The test controls repeatability of placing the internal weight and

determining error of indication with reference to balance’s Max capacity.

Testing Procedure:

Two times loading of the internal weight, followed by 10 times

loading of the internal weight

Balance internal adjustment

Calculating and storing the value of standard deviation

In balances with automatically opened doors, door opening and

closing test is carried out.

Additionally the function enables:

Displaying and printing a report on available printers and saving

the report in balance’s memory. The report contains basic data on

balance’s parameters, ambient conditions and test results.

Test results:

*Deviation for Max.

*Value of repeatability of indication expressed as standard deviation

*Result of door test (positive / negative) – if a balance features

automatic door opening system.