KERN & SOHN AET 200-4NM User Manual

Page 159

Advertising
background image

AET_/PET_/ILT_NM-BA-e-1410

159

Example of a report:

.....................................................

--------- Autotest GLP: Report ---------

Balance type PET

Balance ID 400010

User Admin

Software rev. L0.0.21 S

Date 2012.01.16

Time 09:17:16

----------------------------------------

Number of measurements 10

Reading unit 0.001/0.01 g

Internal weight mass 1402.094 g

Filter Average

Value release Fast & Reliable

----------------------------------------

Deviation for Max. -0.118 g

Repeatability 0.0088 g

Signature

........................................

PROCEDURE:

On pressing a field with test name, the balance displays a message box.

On this level, the user can carry out the following:

Start the following AUTOTESTU GLP process

Preview results of completed autotests, and export all saved test results as

file with extension *.csv

Delete all saved test results

User can abort the process at optional moment by pressing <X> key in the

testing window.

Results of carried out autotests are displayed as a table (each line features date

and time of autotest procedure and its result).

To preview detailed data of an autotest, press a specified line of the table with

test results.

In order to print a result of a single autotest, enter its details (see above) and

generate a printout by pressing <print> key in the upper bar of the balance’s

display.

Autotest results can be exported by pressing <export> field on the level of all

saved reports in a table. Data is sent to a data storage device connected to

balance’s USB port as a file with extension *.csv.

Advertising