KERN & SOHN AET 200-4NM User Manual

Page 30

Advertising
background image

30

AET_/PET_/ILT_NM-BA-e-1410

AUTOTARE

Means of operation:

Each first stable measurement result is tarred. The NET pictogram

flashes on the display. The operator can determine net mass of weighed

load. On taking off the load from balance’s weighing pan, and as the

indication returns to autozero zone, the software automatically

deactivates Tare value.

Automatic footer printout

Accessible options:

MODE - No – manual footer printout,

Sum of measurements – the condition for printing a footer is

exceeding mass value set in parameter <Threshold>

Number of measurements – the condition for printing a footer

is carrying out a pre-defined number of measurements (a

batch) set in parameter <Threshold>

THRESHOLD – setting the value of threshold determining footer

printout.

For option <Sum of measurements> the value is determined

in measuring unit [g], and for option <Number of

measurements> the value is not measurable, and determined

by number of carried out measurements.

Printout mode / printout release

Function key PRINT / printout release, (manual operation)

Never – printout deactivated

First stable – the first stable measurement is printed

Each stable – all stable measurements are set for printing

Each

– printout of all measurements (stable and

unstable), in case of a verified balance only stale

measurement results are printable (as in setting

<Each stable>)

AUTOMATIC MODE

Never – printout deactivated

First stable

– the first stable measurement result obtained after

placing a load on balance’s weighing pan is recorded,

record of the following measurement result takes

place only on unloading the weighing pan, returning

of the mass indication below set threshold’s value,

and placing another load on balance’s weighing pan

Last stable

– accepted is the last stable measurement recorded

before taking the load off the weighing pan. The

record takes place on taking the load off the

balance’s weighing pan, and returning of the mass

indication below set threshold’s value.

THRESHOLD

mass value obligatory for operation of automatic printout. Set in

grams.

Advertising