Minimum sample – KERN & SOHN AET 200-4NM User Manual

Page 31

Advertising
background image

AET_/PET_/ILT_NM-BA-e-1410

31

Printout

Contains type of printout that is related to a working mode. Printout

takes place on pressing PRINT key on balance’s overlay.

Accessible options:

STANDARD PRINTOUT

Possible declaration of printout content: HEADER, WEIGHING DATA

and FOOTER.

printout components marked as <YES> in the menu are printed on

pressing printout activating function key.

NON-STANDARD PRINTOUT

Database of printouts enables selecting one of available non-

standard printouts visible in menu <PRINTOUTS>, or designing a

unique printout which is automatically added to the databases.

Caution: Means of designing printouts is described in point 14.4

of this

user manual.

9.8. Minimum sample

The weighing mode settings comprise a function of <Minimum sample>.

The function is enabled on specifying in the database <Database/Minimum

sample > parameters on methods for determining the minimum sample

and the values of minimum samples specific for a method. In standard

version, this database item is not filled in with data.

Activities related to determining minimum sample and adding data

to the database: <Databases/Minimum samples> are accessible

only for an authorized “KERN” personnel.

Should balance operator use the minimum sample mode, and data on

minimum samples are not entered to balance menu, then please contact

the local representative of “KERN” company.

An authorized KERN representative determines the minimum sample

dependent on packaging mass. The process is carried out with mass

standards, in balance place of use and in accordance with the requirements

of the accepted quality system. The obtained values are entered to balance

software: <Databases/Minimum samples>.

A minimum sample determining method may have a few tare values

defined, along with assigned values of minimum samples, expiry dates of

the carried out measurements and entered data. The settings are not to be

edited by balance user.

Using <Minimum sample> mode guarantees that the results of completed

weighing processes are within set tolerance limits, and they comply with

the requirements of the quality system accepted by an organization.

CAUTION: the mode operates in the weighing mode only.