Caution:

Data on a product, listed in the below table, have to be filled in before

starting a control:

Mass

Nominal mass of a product

Tare

Packaging mass in the adjustment unit

Min

Value of –T1 error (specified in % of nominal mass)

Max

Value of +T1 error (specified in % of nominal mass)

Unit

Measuring unit of a product: [g]

Caution:

Editing product records in the database is described in point 27.2

of this user

manual.

After returning to the main window of the <

SQC> mode press

(control start) soft key located in the bottom bar of the display,

Which automatically opens an editing window <Batch Number> with an on-

screen keyboard (only if in the mode settings of the option <Request batch

number is set to <YES>),

Enter the batch number of controlled batch and accept it by pressing

key.

Caution:

A user can abort the control process in an optional moment by pressing a

(control stop) soft key located in the bottom bar of the display.

On control progress other soft keys and function keys, the

and

keys are blocked.