13.2.3. Readout

The function enables the user to adjust balance operation to ambient

conditions (filter settings) or individual user needs. Menu <Readout> contains

the following elements:

FILTER

Each measurement signal before being displayed is electronically

processed for the purpose of obtaining corrected parameters

specific for stable measurement result, i.e. ready to read.

Balance user can influence to some extend the range of signal processing

by selecting an appropriate FILTER value. Available options:

o

very fast

o

fast

o

average / normal

o

slow

o

very slow

While setting the filtering level the user should consider the actual

operating conditions of a balance. In case of very good operating

conditions set the filter to average/normal or fast, in case of rough

conditions set the filter to slow or very slow.

Caution:

in case of precision balances PET series the recommended range of filter

is between very fast

÷

average/normal;

in case of analytical balances and microbalances the recommended range

of filter is between average/normal

÷

very slow.

Value release

This parameter determines displaying the stability marker for a

measurement result.

There are 3 available settings of value release parameter

o

fast

o

fast + reliable /recommended/

o

reliable

Caution:

The speed of stability marker occurrence depends on applied filter value and

applied value release

Autozero function

The function is to automatically monitor and correct zero

indication of a balance.