KERN & SOHN AET 200-4NM User Manual

Page 27

Advertising
background image

AET_/PET_/ILT_NM-BA-e-1410

27

1. Inserting the specified value of air density and specified value of weighed

sample density to balance’s memory.

On inserting these values the application automatically calculates the

correction factor for weighed mass and indicates corrected value of

weighed mass on balance’s display.
In order to avoid errors, the corrected mass value is accompanied by <

!

>

symbol visible on balance’s display and a printout.

2. The application semi-automatically determines the value of air density. The

specified density of the weighed sample is inserted to balance’s memory by

an operator.

Determination of air density is carried out using a set of two mass

standards. One of them is made of stainless steel, and the other of

aluminum. Based on mass indications of the two standards, the software

automatically calculated the air density and saves the value in balance’s

memory on operator’s confirmation. Then the density of weighed sample

has to be inserted to balance’s memory.

On inserting these values the application automatically calculates the

correction factor for weighed mass and indicates corrected value of

weighed mass on balance’s display.
As in case 1, the corrected mass value is accompanied by <

!

> symbol

visible on balance’s display and a printout.

Air buoyancy correction mode is enabled / disabled from the level of user menu.

The application is operating in weighing mode only.

STARTUP AND OPERATION OF AIR BUOYANCY CORRECTION

PROCEDURE

While in main window press grey coloured workspace

Select option <Settings>

Select parameter <

Air buoyancy correction>

Accessible settings

- Air buoyancy correction – YES/NO

- Sample density (insert the density value of weighed sample). If an

operator uses products from databases, then on product selection

its density value is automatically uploaded.

- Air density – for selecting the means of inserting the air density

value used for calculating buoyancy correction factor.

Advertising