Adding a pipette to the database of pipettes – KERN & SOHN AET 200-4NM User Manual

Page 98

Advertising
background image

98

AET_/PET_/ILT_NM-BA-e-1410

23.3. Adding a pipette to the Database of Pipettes

Database of pipettes contains list of pipettes’ names and other data including

tested volume and errors values for a specific volume. When adding a pipette

to the database, first specify its name, and then add other data on a pipette.

The software operates intuitively, and commands a user to follow displayed

information.
Adding a pipette to the Database can be carried out in menu of the database.

Procedure:

Enter submenu < Databases>, press <

Pipettes> field

Press <

Add > soft key, should a new pipette be added

The software automatically adds a new entry to the database, and enters its

editing mode. Insert all parameters of a new formulation.

List of parameters defined for a pipette:

1. Name: press “Name” field, a window opens for inserting name of a

pipette;

2. Code: for inserting pipette’s code;

3. Model: for inserting model name of a pipette;

4. Tip: for inserting name of used tip;

5. Type of volume: ADJUSTABLE / FIXED;

6. Nominal volume: for specifying nominal volume of a pipette;

7. Minimum volume: for specifying the minimum volume of a pipette (in

case of a pipette with fixed volume, insert <0>);

8. Number of channels: for specifying number of pipette’s channels, in

case of a single channel pipette insert <1>;

9. Type: NONE/A/D1/D2. Pipette type is compatible with applicable

standard. Selecting pipette’s type is mandatory if calibration process

complies with the requirements of standard ISO 8655, as values of

errors differ in each type of a pipette. Therefore, if the software is to

accept adequate error values it is necessary to determine type of a

pipette;

10. Tested volumes: if pressed, a window is opened with a list of

calibrated volumes (in case of a new pipette, the list is empty), add

volume to be tested, and define error value for each tested volume:

Press <

Add > soft key

A window with on-screen numeric keyboard is displayed,

Insert value of tested volume in [µl] and accept by pressing <

> key,

A new record is automatically added to the list with recommended

error values,

In order to change error values press a field with added volume

of a pipette,