AET_/PET_/ILT_NM-BA-e-1410

Switch on continuous transmission in current measuring unit

Format: CU1 CR LF

Accessible responses:

CU1_I CR LF

- command understood but not accessible at this moment

CU1_A CR LF

MASS FRAME

- command understood and in progress

- response is mass value in current measuring unit

Frame format:

1 2 3

4

5

6

7-15

16 17 18 19 20 21

S U I stability

marker space character mass space

unit

CR LF

Switch off continuous transmission in current weighing unit

Format: CU0 CR LF

Accessible responses:

CU0_I CR LF

- command understood but not accessible at this moment

CU0_A CR LF

- command understood and in progress


Set low checkweighing threshold LO

Format: DH_XXXXX CR LF, where XXXXX – mass format

Accessible responses:

DH_OK CR LF - command carried out
ES CR LF

- command not recognized (incorrect mass format)


Set high checkweighing threshold HI

Format: UH_XXXXX CR LF, where XXXXX – mass format

Accessible responses:

UH_OK CR LF

- command carried out

ES CR LF

- command not recognized (incorrect mass format)

Give value of low checkweighing threshold LO

Format: ODH CR LF

Accessible response: DH_MASS CR LF – command carried out

Frame format:

1 2

3

4-12

13

14 15 16

17

18 19

D H

space

mass

space

unit

space

CR LF

Mass - 9 characters with right justification

Unit - 3 characters with right justification