Measuring units – KERN & SOHN AET 200-4NM User Manual

Page 46

Advertising
background image

46

AET_/PET_/ILT_NM-BA-e-1410

If the function is enabled, the following measurement results are

compared to each other in constant time intervals. If the results differ

less than declared AUTOZERO range, e.g. 1 division, the balance is

automatically zeroed, and the markers of stable indication

and

precise zero

are displayed.

Active AUTOZERO means, that each measurement starts from the precise

zero point. There are, however, cases where the function may disturb the

weighing process. For instance during very slow load placing or pouring

onto the balance’s weighing pan. In such case, the correcting system of

zero indication may also correct the actual indication of a load placed on

the weighing pan.

Accessible settings: NO - autozero function disabled

YES - autozero function enabled

Autozero function: Dosing

The function sets autozero operation mode to default one set for

dosing mode.

Accessible values:

NO - autozero operation is automatically disabled on entering

dosing mode

YES - autozero operation is automatically enabled on entering

dosing mode

Last digit

The function determines visibility of the last decimal place

indicated on measurement result. The functions provides three

settings:

Always: all digits are constantly visible

Never: last digit is blanked

When stable: last digit is displayed only on stable measurement

13.2.4. Measuring units

Balance user can declare in a profile: start unit, supplementary unit and two

custom measuring units.

A custom measuring unit features:

o

A multiplier

o

A name (3 characters)

If a custom unit is designed, then its name is added to the list of accessible

measuring units.

This menu additionally enables inserting the value of gravitational acceleration

force for balance’s place of use. It is obligatory should a balance be used to

determine mass in [N].

Advertising