Touch panel calibration, Level control, Touch panel calibration 31.5. level control – KERN & SOHN AET 200-4NM User Manual

Page 154

Advertising
background image

154

AET_/PET_/ILT_NM-BA-e-1410

31.4. Touch panel calibration

Display calibration is required if during operation a user spots incorrect

reaction of the touch panel.

Procedure:

Enter submenu <

Others>

Select parameter <

Touch panel calibration> which opens an editing

window,

Use finger or a thin (but not sharp edged) and soft stylus to press and

hold a display area occupied by a cross. After pressing the fifth cross in

sequence, accept the changes by pressing

key.

Caution:

The process can be aborted by pressing Esc key on an external keyboard

connected to the balance’s terminal.

31.5. Level control

Balance features an Automatic Level Control system. In case of non-verified

instruments, the operation of Automatic Level Control system is user defined.

In case of verified instruments, the settings are hidden and they operate in

accordance with factory settings, i.e. <Active and locked>, weighing process is

possible only if balance is properly leveled.

Procedure:

Enter submenu <

Others>,

Select parameter <

Level control> which opens an editing window,

Select one of available options:

None – level indicator is not displayed, the balance does not control level

status,

Active – level indicator is displayed, the balance alerts level change by

changing colour of level indicator (green → level status OK., red → level

status lost)

Active with lock – level indicator is displayed , the balance alerts level

change by changing colour of level indicator (green → level status OK.,

red → level status lost). If level indicator is red, the display also shows an

alert – no Level - , and blocks weighing function until level status is

restored.