An example of a report from pipette calibration process:

----------- Pipette calibration -----------

User Smith

Client Jones

Pipette p901\1k

Serial no. 7777

No. of channels 1

Channel no. 1

No. of measurements 10

Operation with ISO 8655 Yes

Start date 2012.03.15 07:50:44

End date 2012.03.15 07:54:34

Water temperature 22.15 °C

Temperature 21 °C

Humidity 48 %

Pressure 1005 hPa

Z coefficient 1.00328

------- Tested volume: 1000 µl -------

1 0.998 g 1000.82389 µl

2 0.998 g 1000.82389 µl

3 0.998 g 1000.82389 µl

4 0.998 g 1000.82389 µl

5 0.998 g 1000.82389 µl

6 0.998 g 1000.82389 µl

7 0.998 g 1000.82389 µl

8 0.998 g 1000.82389 µl

9 0.998 g 1000.82389 µl

10 0.998 g 1000.82389 µl

Average volume 1000.82389 µl

Average [%] 100.08 %

Systematic error 0.82389 µl

Systematic error [%] 0.08239 %

Permissible error ± 16 µl

Random error 0 µl

Random error [%] 0 %

Permissible error ± 6 µl

Status Positive

----------------------------------------