KERN & SOHN AET 200-4NM User Manual
Page 112
112
AET_/PET_/ILT_NM-BA-e-1410
Weighing B
Select option:
On selecting the option, the balance moves to the first sample on the list,
for which weighing <B> is not completed. If there are no such samples in a
series, the balance displays a message that such process is inaccessible. If
such process is possible, then new data on initiated process is displayed in
the grey coloured workspace.
Caution:
Description <Weighing B> is accompanied by digits 1/3: <1> means, that
it is the 1 measuring cycle type B, and <3> means, that the <NUMBER OF
REPETITIONS> for this process is set to <3> (3 cycles).
Remember, that the software requests to carry out the <FIRST> cycle out
of <3> from all samples in a series, for which this measurement is
available (a condition is competing Weighing A for the sample).
During process course the bottom bar and the workspace display data on
the following steps that a user should follow.
In case tare is assigned to a sample, then its value is indicated on the
displayed with minus sign.
Place a sample on balance’s weighing pan (if tare is applied, then place a
sample into the container) and press <
> key.
On pressing the above key, a window is displayed for inserting ambient
temperature and humidity value – if option <Ambient conditions> is set to
<VALUE>, or the data on ambient conditions is automatically acquired from
the ambient conditions module and displayed for confirmation – if option
<Ambient conditions> is set to <ONLINE>.
Caution:
If option <Ambient conditions> in the working mode is disabled – set to
<No>, the software skips this part of the process.
On accepting the ambient conditions by pressing <
> key, the software
returns to the mode’s main window and displays a command on unloading
the sample from the weighing pan (displayed in the message bar).
The user should unload the sample from balance’s weighing pan and accept
the activity by pressing <
> key.
The software automatically moves to the following sample in a series.
Follow above procedure and weigh the following samples.
The process can be aborted by pressing <
> key.