AET_/PET_/ILT_NM-BA-e-1410

Weighing B

Select option:

On selecting the option, the balance moves to the first sample on the list,

for which weighing <B> is not completed. If there are no such samples in a

series, the balance displays a message that such process is inaccessible. If

such process is possible, then new data on initiated process is displayed in

the grey coloured workspace.

Caution:

Description <Weighing B> is accompanied by digits 1/3: <1> means, that

it is the 1 measuring cycle type B, and <3> means, that the <NUMBER OF

REPETITIONS> for this process is set to <3> (3 cycles).

Remember, that the software requests to carry out the <FIRST> cycle out

of <3> from all samples in a series, for which this measurement is

available (a condition is competing Weighing A for the sample).

During process course the bottom bar and the workspace display data on

the following steps that a user should follow.

In case tare is assigned to a sample, then its value is indicated on the

displayed with minus sign.

Place a sample on balance’s weighing pan (if tare is applied, then place a
sample into the container) and press <

> key.

On pressing the above key, a window is displayed for inserting ambient

temperature and humidity value – if option <Ambient conditions> is set to

<VALUE>, or the data on ambient conditions is automatically acquired from

the ambient conditions module and displayed for confirmation – if option

<Ambient conditions> is set to <ONLINE>.

Caution:

If option <Ambient conditions> in the working mode is disabled – set to

<No>, the software skips this part of the process.

On accepting the ambient conditions by pressing <

> key, the software

returns to the mode’s main window and displays a command on unloading

the sample from the weighing pan (displayed in the message bar).

The user should unload the sample from balance’s weighing pan and accept
the activity by pressing <

> key.

The software automatically moves to the following sample in a series.

Follow above procedure and weigh the following samples.

The process can be aborted by pressing <

> key.