26.3. Mass control process

Before starting the mass control process carry out the following steps:

Set batch quantity of a carried out mass control process – touch grey

coloured workspace on balance’s display to open a window and select
group <

Settings>. Next select option <

Batch quantity> for

entering quantity of controlled units in a current automatic mass control

process. Accept entered value by pressing

key.

Set <

Low limit> below the value of the smallest (lightest) controlled

part. Set the limit value so that it protects against saving in a final report

mass of a value below the lightest item of the controlled batch, e.g. a half

of a pill.

Set <

High limit> value. Set the limit value so that it protects against

saving in a final report mass of a value exceeding the heaviest item of a

controlled batch, e.g. if two pills are simultaneously dispensed onto a

weighing pan. In such case, the software skips the measurements and

continues mass controlling process.

Set <

Dosing power> parameter defined in [%] and ranging from

0% to 100 %, nominal mass (Mass), limits for errors: negative - <T1->

(Min) and positive - <T1+> (Max). The error values are determined in %

of nominal mass value. The dosing power, mass and error values are

determined for each controlled product separately.

The errors <T2-> and <T2+> are not to be determined by a user. The

software sets their values as doubled value of errors <T1-> and <T1+>.

Setting procedure of control parameters:
when in the main window press SETUP key, and select <

Databases >,

and <

Products> options.

then, select a product entry from the database for which the above values

need to be set. Set required parameters and return to the main menu.

Select a product to be controlled from the database <

>.

On selecting all applicable settings press

key to start automatic mass

control process.

The mass control process is started automatically, the balance is

automatically tarred, and the feeder starts dispensing the first element

onto balance’s weighing pan.

On dispensing the first element onto balance’s weighing pan, the feeder is

stopped and the dispensed item is weighed. The first stable measurement

result is saved.

On saving a measurement result the mass indication on balance’s display

is zeroed and the automatic feeder is restarted to dispense the following

product onto balance’s weighing pan. As the following product is dispensed

onto the weighing pan the weighing process starts. The procedure is

repeated the same way as in case of the first dosed element.

The cycle is repeated until checking (weighing) all of declared pieces in the

<Batch quantity> parameter.