135

List of parameters in a report from pipette calibration process:

1. Pipette [name from the database of pipettes]

2. Serial no. [serial no. inserted during calibration process]

3. Channel no.

4. Start date

5. End date

6. User [name of a user carrying out calibration]

7. Client [client name]

8. Number of measurements

[no. of measurements for each tested

volume]

9. Operation with ISO 8655

[information if errors comply with the

standard]

10. Status [information whether the errors are below permissible

value accepted for the tested volume]

11. Temperature [average temperature value during calibration

process]

12. Humidity [average humidity value during calibration process]

13. Pressure [average pressure value during calibration process]

14. Water temperature [average water temperature during

calibration process]

15. Z coefficient [value if the Z coefficient for given temperature]

27.10. Series

Database of series contains data saved for a series and measurements of

samples.

Procedure

Enter submenu < Databases>, and press <

Series> field

Press <

Add > key to add a new series to the database

If a series already exists in the database, press field with its name to edit

data that is optionally editable


List of parameters defined for a series:

1. Name

2. Code

3. Client

4. Samples

5. Number of samples