KERN & SOHN AET 200-4NM User Manual
Page 166
166
AET_/PET_/ILT_NM-BA-e-1410
Where:
n - weighing platform number
mass - 9 characters with right justification
unit - 3 characters with left justification
Example:
In case there are two weighing platforms.
S I A CR LF – command sent from a computer
P 1 _ ? _ _ _ _ _ _ 1 1 8 . 5 _ g _ _ CR LF
P 2 _ _ _ _ _ _ _ _ _ 3 6 . 2 _ k g _ CR LF - command carried out,
immediate response is mass value from both weighing platforms in basic
measuring unit specific for each of the weighing platforms
Send stable measurement result in current weighing unit
Format: SU CR LF
Accessible responses:
SU_A CR LF
SU_E CR LF
- command understood and in progress
- time limit exceeded while waiting for stable measurement
result
SU_I CR LF
- command understood but not accessible at this moment
SU_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in current measuring unit
Frame format:
1 2
3
4
5
6
7-15
16 17 18 19 20 21
S U space stability
market space character mass space
unit
CR LF
Example:
S U CR LF – command from a computer
S U _ A CR LF – command understood and in progress
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF - command carried out,
response is mass value in current measuring unit.