24.4. An example of process for differential weighing mode

On activating DIFFERENTIAL WEIGHING mode:

Select information to be displayed in the grey coloured workspace

Select appropriate quick access keys

Add a series to balance’s memory (a name identifies a series)

Add samples to a series (a name identifies a sample)

On adding samples return to the main window of the active working

mode

SELECTING a series:

Press quick access key <SERIES>. A

window is displayed with database of

series.

On entering the database, press field

with name of a series. It is automatically

selected for completing.

Name of the selected series is displayed

in the grey coloured workspace (only if

such information is enabled for

displaying in the grey coloured

workspace).

On selecting a series, press one of below listed keys to carry out the

process:

Weighing A

Weighing initial mass of a sample

Tare (T)

Weighing container for a sample - tarring

Weighing (T+A)

Weighing container for a sample (tarring) and weighing a sample

(subsequent processes).

Weighing B

Weighing final mass of a sample. The process is accessible only if a

given series contains samples with completed weighing records

<A>. unless such weighing records are in series, the process is

unavailable.