Manual printout / automatic printout – KERN & SOHN AET 200-4NM User Manual

Page 170

Advertising
background image

170

AET_/PET_/ILT_NM-BA-e-1410

33.3. Manual printout / Automatic printout

Balance enables generating manual or automatic printouts.

Manual printout: press

key on stabilization of indication (measurement

result),

Automatic printout is generated automatically, in accordance to the settings

of automatic printout (see point 10.8 of this user manual).

The content of a printout depends on setting in menu <Standard printout> -

<Weighing printout template> (see point 15.5 of this user manual).

Format of mass printout:

1

2

3

4 -12

13

14 15 16 17 18

stability

marker

space character mass space

unit

CR LF

Stability marker

[space] if stable

[?] if unstable

[!] air buoyancy correction mode is enabled

[^] if error of exceeding range to + occurs

[v] if error of exceeding range to – occurs

Character

[space] for positive values

[-] for negative values

Mass

9 characters with decimal point and right justification

Unit

3 characters with left justification

Example:
_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF
– a printout generated from a balance

on pressing

key, according to settings of <Weighing printout template>:

N (number of
measurements)

NO

Universal variable 1 … 5

NO

Date

NO

Net

NO

Time

NO

Tare

NO

Level status

NO

Gross

NO

Client

NO

Current result

NO

Warehouse

NO

Supplementary unit

NO

Product

NO

Mass

YES

Packaging

NO

Non-standard printout

NO


Advertising