26. MASS CONTROL

Working mode < Mass control > enables controlling mass of a product that is

automatically batched (dosed) onto the balance’s weighing pan by an automatic

feeder PA-02/H that is connected to the balance. The working mode

automatically controls a complete batch of tested product after determining its

quantity.

26.1. Working mode activating procedure

Procedure:

while in the main window, press

soft key located in the upper bar of the

display, which opens a submenu <Working modes> with selection of

available working modes

select < Mass control > the software automatically returns to the main
window and displays

pictogram in the upper bar,

Automatically a single on-screen button appears in the bottom bar of the

display:

Start mass control proccess

26.2. Mass control global settings

Entering the mode settings is carried out by pressing the grey coloured

workspace and selecting an group of parameters named "Settings" <

>,

which contains the following:

Request batch

number

The function orders requesting a batch

number before starting mass control

process.

Batch quantity

Determines quantity of a batch, i.e. number

of product units to be controlled.

Low limit

A value exceeding which causes accepting a

measurement record by a balance.

High limit

A value exceeding which causes omitting a

measurement in the summary, e.g. if two

parts (pills) are simultaneously dispensed on

balance’s weighing pan.

Print report

A parameter enabling for switching on/off

automatic printout of a report on completing

mass control process.

Printout

A parameter enabling selecting whether on

completing the control process the balance

should trigger a standard or a non-standard

printout.