Working mode activating procedure – KERN & SOHN AET 200-4NM User Manual

Page 118

Advertising
background image

118

AET_/PET_/ILT_NM-BA-e-1410

25.1. Working mode activating procedure

Procedure:

while in the main window, press

soft key located in the upper bar of the

display, which opens a submenu <Working modes> with selection of

available working modes,

select <

SQC>, mode, the software automatically returns to the main

window and displays

pictogram in the upper bar,

Simultaneously, the message bar displays a command <Start control > and

a soft-key on control start is visible in the bottom tool bar:

Start control

Control start procedure:

Starting a control process requires logging a user with corresponding access

level.

Caution:

1.

Starting a control requires that an access level of the logged user is at

least a <user>. If a logged user or an anonymous user have access level

set to <guest> then on control start the software opens a message box

informing on: <Unauthorized access>.

2.

Logging procedure is described in point 9 of this user manual, and

procedure determining access level for each user is provided in point 13 of

this user manual.

Set general settings of working mode parameters to balance memory
<

SQC>, description in point 25,

Select a needed product (by pressing a corresponding soft key
<

> in the bottom bar or using balance’s keyboard with correctly determined

data on a control: