31.6. Sensor sensitivity

The parameter is adjustable between 0 – 9, and it determines the

distance from which the IR proximity sensors should react.

As standard, the sensitivity is set within the range of 5 to 7.

Procedure:

Enter submenu <

Others>

Select parameter <Sensor sensitivity> which opens an editing window,

Select one of available options. The selection causes immediate returning

to the menu window.


31.7. Autotest

AUTOTEST function is designed to aid a user in both assessing balance’s

operation and diagnosing the reasons for occurrence of errors in weighing

which exceed the maximum permissible values for a given balance

model.

By means of a simple, repeatable and fully documented way the function

enables optimizing balance’s settings to maintain the best possible

repeatability and weighing time at a workstation. The main purpose of

function operation is the possibility of monitoring the above mentioned

parameters at optional moment and saving records from carried out

tests.

The functions is divided into two modules:

AUTOTEST FILTER

AUTOTEST GLP

Before carrying out any of the tests , the balance controls its level status,

temperature and humidity.