9.3. Balance leveling

Balance leveling procedure

Press level status pictogram <

> located in the upper right corner of the

display.

The display indicates a control panel of leveling function. Balance operator

previews level status, and balance’s top elevation.

Level the balance by turning the adjustable feet in a way indicated by
blinking pictograms on the display < > – the level point should move

towards the center of the level circle.

As the level point is in the center point, its colour changes from red to

green – which is a confirmation for correct level status.

9.4. Zeroing

In order to zero mass indication press

key.

The mass indication on the display should equal zero, and precise zero

and

stability

markers should appear.

Zeroing process is tantamount to determining new zero point recognized by

the balance as precise zero. Zeroing is possible only in stable status of display

indication.

Caution:

Zeroing the display indication is possible only within the ±2% range of

instrument’s maximal capacity. If zeroed value is above the ±2% of the

maximal capacity, then the software indicates an error message Err2.