Tarring, Use of air buoyancy correction factor, Tarring 9.6. use of air buoyancy correction factor – KERN & SOHN AET 200-4NM User Manual

Page 25

Advertising
background image

AET_/PET_/ILT_NM-BA-e-1410

25

9.5. Tarring

In order to determine net weight of a weighed object, place object’s container

(packaging) on balance’s weighing pan, and on stabilization of measurement

result press

key. The display should indicate mass equal to zero and

symbols: Net and

. On taking off the weighed load and its packaging from

instrument’s weighing pan, the display indicates sum of total tarred mass with

minus sign.

The software also enables assigning tare value to a product from a database.

Then, on product selection from a database, the software automatically

uploads data on tare value for the specific product.

Caution:

Tarring negative values is impossible. On tarring negative values the balance

responds with an error message. In such case zero balance’s indication and repeat

tarring procedure.

Manual tare determination

Procedure:

While in optional mode press quick access key

,

Which opens numeric keyboard on balance’s display,

Using numeric keys insert desired tare mass and press

key,

The balance returns to weighing mode, and the display indicates inserted

tare value with minus “–“ sign.

Deleting tare

The tare value indicated on balance’s display can be deleted by pressing ZERO

key on balance’s front panel, or using programmable function key <Deactivate

tare>

PROCEDURE 1 – on taking the tarred load off balance's weighing pan

Press ZERO key

The NET marker is deleted, and new zero point of the balance is determined

PROCEDURE 2 – when the tarred load in on balance’s weighing pan

Press ZERO key

The NET marker is deleted, and new zero point of the balance is determined

If tare value exceeds 2% of balance’s maximal capacity, the display

indicates error message –Err- (forbidden operation)

PROCEDURE 3 - when the tarred load in on balance’s weighing pan or on

taking the tarred load off balance's weighing pan

Press programmable key <Deactivate tare>

The NET marker is deleted,

The display indicates tare value

Press <Restore tare>

key to restore the last tare value

9.6. Use of air buoyancy correction factor

Advertising