Adjustment, Internal adjustment, External adjustment – KERN & SOHN AET 200-4NM User Manual

Page 35: Internal adjustment 10.2. external adjustment

Advertising
background image

AET_/PET_/ILT_NM-BA-e-1410

35

10. ADJUSTMENT

The Balances feature automatic internal adjustment system which ensures

correct measurement accuracy. Menu <ADJUSTMENT> contains functions

controlling operation of balance adjustment process, including options:

10.1. Internal adjustment

Internal adjustment process utilizes an internal weight built in balance’s

housing. <Internal adjustment> function key activates automatic adjustment

process. On its completion, balance’s display indicates a message box on

process completion and its status.

CAUTION:

Balance adjustment requires stable measurement conditions (free from air breeze

and vibrations), adjustment process has to be carried out with empty weighing pan.

10.2. External adjustment

External adjustment is carried out using an external adjustment weight, with

appropriate accuracy and mass relating to balance’s maximal capacity and

readability. The process takes place semi-automatically, and the following

process phases are indicated on balance’s display.

CAUTION:

External adjustment is available only in balances which are not subject to conformity

assessment (verification).

Process course:

Enter submenu < Adjustment > and select option: “ External

adjustment”,

Balance’s display indicates the following message box

If there is a load on balance’s weighing pan, unload it,

Press

key. The balance determines start mass, which is indicated by

a message box: “Start mass determination” on balance’s display

On determining the start mass, the balance display’s a message box:

Advertising
This manual is related to the following products: