Additional settings of density mode – KERN & SOHN AET 200-4NM User Manual

Page 74

Advertising
background image

74

AET_/PET_/ILT_NM-BA-e-1410

9. Load aluminum mass standard on balance’s weighing pan and on

stabilization of measurement result accept it by pressing <

> key,

10. The balance automatically calculates the density of air, and indicates its

value on the display,

11. Press <

> key to finish the procedure.

The value of determined air density is automatically assigned to field <Air

density> in menu <Settings/ Air buoyancy compensation> in working mode:

weighing.

19.4. Additional settings of Density mode

The additional settings enable adjusting the working mode to user’s needs and

requirements. To access the setting follow below procedure:

Procedure:

1. Press grey coloured workspace,

2. The display indicates menu of: Settings, Keys, Information, Printout

and Profile,

3. Press <Settings> key,

4. The display indicates functions related to density determining

process, such as:

o

Request sample number

Available settings:

NO – data on sample number is not required; applicable in case of

a single measurement

YES – before each measurement a window requesting inserting

sample number is automatically displayed; applicable in case of a

series of measurements

o

Measuring unit: the balance enables choosing between the

following units: [g/cm

3

], [kg/m

3

], [g/l]. Selected unit is activated

for all density determining options and also in final printout.