23. PIPETTE CALIBRATION

Working mode <

Pipette calibration> is dedicated to determining

measurement errors of volume in piston pipettes, in accordance with standard

ISO 8655 or according to user requirements.

In case of calibration carried out in accordance with standard ISO 8655, the

errors are automatically selected to tested volume, as specified in the standard

(see table of errors for standard ISO 8655).

Working mode enables:

Calibrating pipettes with fixed or adjustable volume, single or multi-

channel;

Preparing a database of pipettes including: pipette name, code, tested

volume and others;

Calculating results on basis of:

Average volume of a pipette (channel)

Systematic error es (accuracy error)

Random error CV (repeatability error)

Carrying out automated measuring procedures with consideration of

pipette’s type;

Storing calibration results in a database (in a form of reports from

calibration process);

Printing reports from pipette calibration process

Exporting reports from tests

During volume testing procedure, the software determines accuracy and

repeatability errors for tested volume.

In case of pipettes with adjustable volume, the software enables declaring up to

5 volume from total pipette’s range, that are checked during calibration process.

In order to ensure the highest accuracy of pipettes calibration process,

maintain the following ambient conditions at a workstation:

Ambient temperature of a pipette, tips and liquid should be kept between

20°C - 25°C with change rate during testing within ± 0.5°C

Relative humidity 50 - 75%

and

Use distilled water for pipettes calibration processes

A pipette, tips and distilled water should be stabilized for temperature in

the weighing room. The reference norm advises that minimum

acclimatization time for above mentioned is 2 hours.

While calibrating process, balance user refers to the database of pipettes

including data on pipettes, its parameters, tested volume, and values of errors

determined for specific volumes.

Before carrying out calibration process, load database with data on pipettes and

calibration criteria.