A5.5 output process, A5.5.1 status determination, A5.5 output – Yokogawa digitalYEWFLO (DY-FF) User Manual

Page 102: Process, A5-6, A5.5.1 status, Determination, Good uncertain bad

Advertising
background image

A5.5 Output Process

There are the following three output parameters:

1. OUT
2. OUT_TRIP
3. OUT_PTRIP

Parameters OUT_TRIP and OUT_PTRIP are used
only when INTEG_TYPE is a value from 1 to 4.
In case of Integrator block related memory failed,
the status of OUT, OUT_TRIP, OUT_PTRIP
becomes “Bad-Device Failure”.

OUT.Value, OUT_TRIP.Status, and OUT_PTRIP.
Status are determined by the ratio of the “Good”
integrated values to all integrated values, which is
stored in PCT_INCL (0% to 100%). The user must
set the threshold value of each status to UNCERT_
LIM and GOOD_LIM.
The Integrator block determines the status of the
output using the three parameters: PCT_INCL,
UNCERT_LIM, and GOOD_LIM.
● PCT_INCL≥GOOD_LIM

Good

● UNCERT_LIM≤PCT_INCL<GOOD_LIM

Uncertain

● PCT_INCL<UNCERT_LIM

Bad

If INTEG_TYPE is 5, 6, or 7, the status of the trip
output becomes “Good-NS-Constant.”

0%

UNCERT_LIM

GOOD_LIM

100%

PCT_INCL

GOOD

Uncertain

Bad

PCT_INCL=100

×(1 - (msp of RTotal)/(msp of ATotal))

msp of RTotal: RTotal value that is converted into a short floating-point number
msp of ATotal: ATotal value that is converted into a short floating-point number
RTotal: Integrated value of the absolute values of the increments whose status is bad
ATotal: Integrated value of the absolute values of the increments regardless of the output status

FA0504.ai

Figure A5.4 Status of OUT, OUT_TRIP, and OUT_PTRIP Outputs

A5.5.1 Status

Determination

The same criteria for determining the status of the
output of the Integrator block are used in common
for the above three parameters.

<APPENDIX 5. INTEGRATOR (IT) BLOCK>

A5-6

IM 01F06F00-01EN

Advertising